本文へスキップ

AIST NMIJ/Material Strength Standards Group|Japanese|

Calibration Services

Our group provides measurement standards related to the hardness field and underwater ultrasonic waves.
For inquiries and applications for calibration services, please visit the Calibration and Contract Testing page of the National Metrology Institute.

Hardness standard

■Rockwell hardness standard

Rockwell hardness test is measurement method capable of direct numerical reading, and widely-used method in industrial field such as automobile. NMIJ provides Rockwell C scale hardness standard (for steel material), and Rockwell B scale hardness standard (for soft iron, brass).

Calibration type Rockwell C scale hardness
Calibration item Rockwell hardness standard block
(C scale)
Scope of calibration 20 HRC-65 HRC

Calibration type Rockwell B scale hardness
Calibration item Rockwell hardness standard block
(B scale)
Scope of calibration 30 HRBW-100 HRBW
Note Provided only with a tungsten carbide ball indenter

■Vickers hardness standard

Vickers hardness test is applied to samples of various material and thickness due to less dependence in test force. NMIJ provides Vickers hardness standard in the range of 200 HV-900 HV.

Calibration type Vickers hardness
Calibration item Vickers hardness standard block
Scope of calibration 200 HV-900 HV

■Brinell hardness standard

Brinell hardness test can obtain average hardess with respect to relatively large area. NMIJ provides Brinell hardness standard in the range of 200 HBW-500 HBW (for mainly soft steel).

Calibration type Brinell hardness
Calibration item Brinell hardness standard block
Scope of calibration 200 HBW-500 HBW
Note Provided only with a tungsten carbide ball indenter
Test force symbol: HBW2.5/187.5-HBW10/3000


Ultrasonic


■Ultrasonic power standard

Ultrasonic power is total energy of ultrasound emitted from an ultrasonic transducer per unit time and is one of the key quantities for evaluating ultrasonic field, as well as ultrasonic pressure. We have been providing the calibration service for ultrasonic power in the range of 0.1 mW to 100 W using the following two methods. In the range below 15 W, ultrasonic power is calibrated using radiation force balance method using an electronic balance. In the range above 15 W, it is calibrated using calorimetric method, which measures the ultrasonic power from the heat generation of water.

Device under test Circular ultrasonic transducer
Calibration range 1 mW ≤ P ≤ 15 W (0.5 MHz ≤ f ≤ 15 MHz)
1 mW ≤ P ≤ 500 mW (15 MHz ≤ f ≤ 20 MHz)
15 W ≤ P ≤ 100 W (1 MHz ≤ f ≤ 3 MHz)
Calibration and
measurement
capability
5 % - 12 % (1 mW ≤ P ≤ 15 W)
9 % (15 W ≤ P ≤ 100 W)
Calibration method Radiation force balance method
(based on IEC 61161)
and calorimetric method


■Ultrasonic pressure standard (Calibration of free-field sensitivity for a hydrophone)

A hydrophone is an underwater device that transforms ultrasonic pressure into voltage signal. A hydrophone sensitivity, a coefficient convert voltage signal into ultrasonic pressure, is required in order to measure ultrasonic pressure using a hydrophone precisely. We have been providing the calibration service for hydrophone sensitivity by substitution method using the reference hydrophone calibrated by reciprocity and interferometric methods.

Device under test Hydrophone
Calibration range 0.1 MHz ≤ f ≤ 40 MHz
Calibration and
measurement
capability
10 % - 13 % (0.1 MHz ≤ f ≤ 1 MHz)
6.1 % - 8.8 % (0.5 MHz ≤ f ≤ 20 MHz)
13 % (21 MHz ≤ f ≤ 30 MHz)
17 % (31 MHz ≤ f ≤ 40 MHz)
Calibration method Substitution calibration using the reference
hydrophone calibrated by reciprocity method
(based on IEC 60565, range: 0.1 MHz-1 MHz)
and interferometric method
(based on IEC 62127-2, range: 0.5 MHz-40 MHz)