The water flow calibration facilities cover the wide range between 2 L/h and 12 000 m3/h with uncertainties between 0.039 % and 0.081 %.
■Hi-Reff (High Reynolds number actual flow facility)
General Features
Flow Rate |
750 ~ 12 000 m3/h |
Uncertainty |
0.081 % (k=2) |
Calibration Method |
Master Meter Method |
Temperature |
~ 70 °C |
■Medium and Large
General Features
Flow Rate |
0.3 ~ 3 000 m3/h |
Uncertainty |
0.042 ~ 0.060 % (k=2)
|
Calibration Method |
Static and Gravimetric Method with Flying Start and Stop Method |
■Small
General Features
Flow Rate |
0.002 ~ 1.2 m3/h |
Uncertainty |
0.04 % (k=2)
|
Calibration Method |
Static and Gravimetric Method with Flying Start and Stop Method |